4.4 Article

Intra-session repeatability of quantitative metrics using widefield optical coherence tomography angiography (OCTA) in elderly subjects

期刊

ACTA OPHTHALMOLOGICA
卷 98, 期 5, 页码 E570-E578

出版社

WILEY
DOI: 10.1111/aos.14327

关键词

filters; macula; optical coherence tomography angiography; perfusion density; repeatability; vessel density

资金

  1. National Medical Research Council [NMRC/CG/C010A/2017]
  2. SingHealth Medical Student Talent Development Award-Travel (SMSTDA-Travel)
  3. Duke-NUS Medical School [Duke-NUS-KP(Coll)/2018/0009A]

向作者/读者索取更多资源

Purpose To assess the repeatability of retinal vascular metrics using different postprocessing methods as obtained from the swept-source optical coherence tomography angiography (SS-OCTA). Methods Thirty-two participants (63% males; mean [SD] age, 70 [7] years) underwent SS-OCTA imaging (PLEX (R) Elite 9000, Carl Zeiss Meditec, Inc., Dublin, USA). Each participant underwent 2 repeated scans of 2 scan protocols: a macular-centred 3 x 3-mm(2) and a widefield 12 x 12-mm(2) for a total of 4 acquisitions. Images of superficial vascular plexuses (SVP) and deep vascular plexuses (DVP) were processed using different filters to generate the perfusion density (PD) and vessel density (VD). Vessel enhancement filters ranged from vessel targeted (Hessian and Gabor filters), classical denoising (Gaussian filter), to a scale-selective adaption (modified Bayesian residual transform [MBRT]). Intra-session repeatability of the different filters and their correlation with the original data set were calculated with the intraclass correlation coefficient (ICC) and Pearson's r. Results Of the 32 eyes, 17 and 15 were right and left eyes, respectively. For 3 x 3-mm(2) scans, both MBRT and Gabor filters yielded very good repeatable PD and VD (both ICCs > 0.87) values. Gabor filter was the most correlated with the original data set for the OCTA metrics (r = 0.95-0.97). For 12 x 12-mm(2) scans, MBRT filter produced good-to-moderate ICC values for SVP (ICC>0.89) and DVP (ICC>0.73) metrics. Both the MBRT and Gabor filters were highly correlated with the original 12 x 12-mm(2) scan data set (r = 0.96-0.98). The ICCs for the agreement between 3 x 3-mm(2) and cropped 12 x 12-mm(2) were high only for the PD values at the SVP layer and were poor for the VD at SVP and DVP measurements (ICC < 0.50). Conclusion Our findings show that with the proper choice of postimaging processing methods, SS-OCTA metrics can be obtained with high repeatability, which supports its use in various clinical settings.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据