Validation of Omron HBP-1300 professional blood pressure monitor based on auscultation in children and adults

期刊

BMC CARDIOVASCULAR DISORDERS
卷 16, 期 -, 页码 -

出版社

BIOMED CENTRAL LTD
DOI: 10.1186/s12872-015-0177-z

关键词

Validation study; Electronics; Sphygmomanometer

资金

  1. Guangzhou Boji Medical Biotechnological Co. Ltd [HBP-1300]

向作者/读者索取更多资源

Backgroud: To determine whether the professional Omron HBP-1300 blood pressure (BP) monitor meets American Association for the Advancement of Medical Instrumentation (AAMI) accuracy standards in Chinese children and adults. Method: According to the AAMI protocol, simultaneous auscultatory measurements by two observers using a mercury manometer were obtained in participants using the Omron HBP-1300. Triple measurements were obtained after a minimum 5-min rest with a 1-min interval between adjacent measurements. Results: A total of 85 participants submitted to 255 doctor-measured BP and 255 successful professional monitor readings. The initial auscultation systolic BP was <100 mmHg in 25 participants (29.4 %), 100-160 mmHg in 53 participants (62.4 %), and >160 mmHg in seven participants (8.2 %). All of the simultaneous measurements agreed to within +/- 10 mmHg, while 95 % agreed to within +/- 4 mmHg for both systolic and diastolic BP, and the consistency between two observers was satisfactory. The difference between the devices was -1.3 +/- 3.6 mmHg for systolic BP and 0.7 +/- 3.8 mmHg for diastolic BP and by AAMI method 1, which met this guideline. The average difference between two devices by AAMI method 2 was 1.4 +/- 3.2 mmHg for systolic BP and 1.0 +/- 3.9 mmHg for diastolic BP, which met this guideline. Conclusion: The professional BP monitor Omron HBP-1300 is desirable for measuring the BP for Chinese children and adults.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据