Repeatability and comparison of new Corvis ST parameters in normal and keratoconus eyes

期刊

SCIENTIFIC REPORTS
卷 9, 期 -, 页码 -

出版社

NATURE RESEARCH
DOI: 10.1038/s41598-019-51502-4

关键词

-

资金

  1. Open Program of Shandong Provincial Key Laboratory of Ophthalmology [2018-04]
  2. Henan Natural Science Foundation [182300410362]
  3. Henan Key Laboratory of Ophthalmology and Visual Science
  4. National Key Clinical Specialties Construction Program of China

向作者/读者索取更多资源

To evaluate the repeatability of corneal biomechanical parameters in normal and keratoconus eyes, and explore factors that affects the repeatability, and further assess the diagnostic ability of new parameters. Seventy-seven keratoconus eyes of 47 patients and 77 right eyes of 77 normal subjects were recruited in current study. All participants received three repeated measurements with 2 to 5 minutes interval. The interclass correlation coefficient (ICC), Cronbach' alpha and repeatability coefficient (RC) were evaluated. The liner regression analysis was used to identify factors that affect the repeatability, and linear mixed effects model was performed to compare the parameters differences. The receiver operating characteristic (ROC) curve was used to evaluate the diagnostic ability of new parameters. Eighteen parameters in normal eyes and twenty-two parameters in keratoconus eyes showed excellent repeatability (ICC >= 0.90). Age, axial measurement (AL), spherical equivalent, astigmatism, gender, mean keratometry (Kmean), intraocular pressure (IOP) and central corneal thickness (CCT) could affect the repeatability of new Corvis ST parameters. Compared with normal eyes, the Ambrosio's Relational Thickness horizontal (ARTh), biomechanical corrected IOP (bIOP), stiffness parameter at first applanation (SP A1) were low and the Max Inverse Radius, deformation amplitude (DA) Ratio Max [2 mm], Pachy Slope, DA Ratio Max [1 mm], Integrated Radius and Corvis Biomechanical Index (CBI) were high in keratoconus eyes (All P < 0.05). Both ARTh and CBI had high Youden index (0.870), and the corresponding cut-off values were 379.29 and 0.44. The repeatability of Corvis ST parameters was acceptable both in normal and keratoconus eyes, and new parameters could effectively diagnose keratoconus eyes from normal eyes.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据