期刊
BIOMEDICAL OPTICS EXPRESS
卷 10, 期 12, 页码 6462-6475出版社
OPTICAL SOC AMER
DOI: 10.1364/BOE.10.006462
关键词
-
资金
- Boehringer Ingelheim Stiftung, Mainz/Germany
- International PhD Programme on Gene Regulation, Epigenetics & Genome Stability, Mainz/Germany
Single molecule localization microscopy (SMLM) has been established to acquire images with unprecedented resolution down to several nanometers. A typical time scale for image acquisition is several minutes to hours. Yet it is difficult to avoid completely sample drift for long time measurements. To estimate drift, we present a method based on the evaluation of speckle patterns formed by backscattered laser light from the cells using a single molecule localization microscope setup. A z-stack of unique speckle patterns is recorded prior to the measurements as a three-dimensional position reference. During the experiment, images of scattered laser light were acquired, and correlated individually with each of the images of the speckle reference stack to estimate x, y and z drift. Our method shows highly comparable results with a fiducial marker approach, achieving a precision of several nanometers. This method allows for high precision three dimensional drift correction of microscope systems without any additional sample preparation. (C) 2019 Optical Society of America under the terms of the OSA Open Access Publishing Agreement
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐