4.3 Article

Validation of methods for determining pediatric midazolam using wet whole blood and volumetric absorptive microsampling

期刊

BIOANALYSIS
卷 11, 期 19, 页码 1737-1754

出版社

FUTURE SCI LTD
DOI: 10.4155/bio-2019-0190

关键词

blood; hematocrit; LC-MS; microsampling; Mitra R; microsamplers; pediatric; pharmacokinetics; volumetric absorptive sampling (VAMS); wet vs dry

向作者/读者索取更多资源

Aim: Collection and quantitative analysis in dry blood using volumetric absorptive microsampling (VAMST) potentially offers significant advantages over conventional wet whole blood analysis. This manuscript explores their use for pediatric sampling and explores additional considerations for the validation of the bioanalytical method. Results: HPLC-MS/MS methods for the determination of midazolam and its major metabolite 1-OH midazolam in both whole wet blood, and dry blood collected on VAMS were developed, validated, and used to support an observational clinical study to compare pharmacokinetic parameters in pediatric patients. Conclusion: Validation data met internationally accepted guideline criteria. A strong correlation was observed in calculated concentrations between wet and dry test samples, indicating that VAMS is a suitable technique for use in pediatric clinical studies.