4.7 Article

Fast quantification of whisky lactone in oak wood by ion mobility spectrometer

期刊

TALANTA
卷 209, 期 -, 页码 -

出版社

ELSEVIER
DOI: 10.1016/j.talanta.2019.120567

关键词

Whisky lactone; Ion mobility spectrometry; ims; Atmospheric pressure chemical ionisation; apci; Mass spectrometry

资金

  1. European Union [692335, 674911]
  2. Slovak Research and Development Agency [APVV-17-0318, APVV-15-0580]
  3. Slovak Grant Agency for Science [VEGA 1/0733/17]
  4. MEPhI Academic Excellence Project

向作者/读者索取更多资源

An Ion Mobility Spectrometry (IMS) apparatus has been used to detect the beta-methyl-gamma-octalactone (Whisky Lactone - WL) standard in the air and as well WL vapours originating from the oak wood samples. The IMS was equipped with Atmospheric Pressure Chemical Ionisation (APCI) ion source based on a Corona Discharge (CD) and was operated in the positive polarity. The IMS spectrum of WL exhibits two peaks, a monomer with reduced ion mobility value K-0 = 1.39 cm(2 )V(-1) s(-1) and dimer K-0 = 1.09 cm(2) V-1 s(-1). Using Ion Mobility orthogonal acceleration Time of Flight mass spectrometer (IMS-oaTOF MS) these peaks were identified as protonated monomer M center dot H+center dot(H2O)(0,1) and dimer M-2 center dot H+ ions respectively. The limit of detection study resulted in LOD for WL of 50 PPbv. Detection of WL from oak wood samples of different Quality Level (QL) categories (categories 1 to 10), indicated strong correlation between the QL category number and the response of the IMS.