4.8 Article

Integrated microfluidic device using a single universal aptamer to detect multiple types of influenza viruses

期刊

BIOSENSORS & BIOELECTRONICS
卷 86, 期 -, 页码 247-254

出版社

ELSEVIER ADVANCED TECHNOLOGY
DOI: 10.1016/j.bios.2016.06.071

关键词

Aptamer; Influenza virus; Magnetic beads; Microfluidics

资金

  1. Ministry of Science and Technology of Taiwan (MOST) [104-2218-E-007-023, 104-2119-M-007-009]
  2. National Health Research Institute of Taiwan [NHREI-EX104-10428EI]
  3. Towards a World-Class University Project from National Tsing Hua University

向作者/读者索取更多资源

DNA aptamers that can bind specific molecular targets have great potential as probes for microbial diagnostic applications. However, aptamers may change their conformation under different operating conditions, thus affecting their affinity and specificity towards the target molecules. In this study, a new integrated microfluidic system was developed that exploited the predictable change in conformation of a single universal influenza aptamer exposed to differing ion concentrations in order to detect multiple types of the influenza virus. Furthermore, the fluorescent-labeled universal aptamer used in this system could distinguish and detect three different influenza viruses (influenza A H1N1, H3N2, and influenza B) at the same time in 20 min and therefore has great potential for point-of-care applications requiring rapid diagnosis of influenza viruses. (C) 2016 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据