Validation of a Clinical Aberrometer Using Pyramidal Wavefront Sensing

期刊

OPTOMETRY AND VISION SCIENCE
卷 96, 期 10, 页码 733-744

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/OPX.0000000000001435

关键词

-

资金

  1. CooperVision, CA

向作者/读者索取更多资源

PURPOSE This study examines the accuracy and repeatability of a newly available high-resolution pyramidal wavefront sensor-based aberrometer (Osiris by Costruzione Strumenti Oftalmici, Firenze, Italy). METHODS An engineered model eye and a dilated presbyopic eye were used to assess accuracy and repeatability of aberration measurements after systematic introduction of lower- and higher-order aberrations with calibrated trial lenses (sphere +10.00 to -10.00 D, and astigmatic -4.00 and -2.00 D with axis 180, 90, and 45 degrees) and phase plates (-0.57 to 0.60 mu m of Seidel spherical aberration defined over a 6-mm pupil diameter). Osiris aberration measurements were compared with those acquired on a previously calibrated COAS-HD aberrometer for foveal and peripheral optics both with and without multizone dual-focus contact lenses. The impact of simulated axial and lateral misalignment was evaluated. RESULTS Root-mean-square errors for paraxial sphere (corneal plane), cylinder, and axis were, respectively, 0.07, 0.11 D, and 1.8 degrees for the engineered model and 0.15, 0.26 D, and 2.7 degrees for the presbyopic eye. Repeatability estimates (i.e., standard deviation of 10 repeat measures) for the model and presbyopic eyes were 0.026 and 0.039 D for spherical error. Root-mean-square errors of 0.01 and 0.02 mu m, respectively, were observed for primary spherical aberration and horizontal coma (model eye). Foveal and peripheral measures of higher- and lower-order aberrations measured with the Osiris closely matched parallel data collected with the COAS-HD aberrometer both with and without dual-focus zonal bifocal contact lenses. Operator errors of focus and alignment introduced changes of 0.018 and 0.02 D/mm in sphere estimates. CONCLUSIONS The newly available clinical pyramidal aberrometer provided accurate and repeatable measures of lower- and higher-order aberrations, even in the challenging but clinically important cases of peripheral retina and multifocal optics.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据