4.5 Article

Experimental and Monte Carlo determination of HPGe detector efficiency

期刊

出版社

SPRINGER
DOI: 10.1007/s10967-019-06856-4

关键词

HPGe; Monte Carlo; Geant; Full energy peak efficiency

资金

  1. Slovak Research and Development Agency [APVV-15-0017]

向作者/读者索取更多资源

Monte Carlo model has been developed using the GEANT code for calculation of the full energy peak efficiency of a HPGe detector operating in the Slovak Institute of Metrology. The model has been used for calculation of the HPGe detector efficiency for gamma-spectrometry measurements associated with the development of a national radon standard. The detector model was validated by comparison of simulated efficiencies with measured experimental values for point sources and for 450 mL Marinelli beaker. A reasonable, up to 5% agreement between the simulated and experimental results was achieved in the energy range from 100 to 1800 keV.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据