JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY
卷 322, 期 3, 页码 1863-1869出版社
SPRINGER
DOI: 10.1007/s10967-019-06856-4
关键词
HPGe; Monte Carlo; Geant; Full energy peak efficiency
资金
- Slovak Research and Development Agency [APVV-15-0017]
Monte Carlo model has been developed using the GEANT code for calculation of the full energy peak efficiency of a HPGe detector operating in the Slovak Institute of Metrology. The model has been used for calculation of the HPGe detector efficiency for gamma-spectrometry measurements associated with the development of a national radon standard. The detector model was validated by comparison of simulated efficiencies with measured experimental values for point sources and for 450 mL Marinelli beaker. A reasonable, up to 5% agreement between the simulated and experimental results was achieved in the energy range from 100 to 1800 keV.