Antegrade versus retrograde screw fixation of anterior column acetabular fractures: a biomechanical in vitro study

期刊

出版社

SPRINGER HEIDELBERG
DOI: 10.1007/s00068-019-01255-6

关键词

Acetabular fracture; Acetabulum; Pelvis; Osteoporosis; Biomechanical; Screw fixation

资金

  1. Stephan Perren AOTrauma Research Traveling Fellowship

向作者/读者索取更多资源

In this in vitro mechanical study, antegrade and retrograde screw fixation of anterior column acetabular fractures showed no significant differences in construct survival, load to failure, stiffness, and mode of failure, with transiliac fracture as the main mode of failure in all samples.
Background To compare the mechanical strength of antegrade versus retrograde lag screw fixation of anterior column acetabular fractures. Methods Standardised anterior column fractures were created in synthetic pelvis models and stabilised by either antegrade (ANTE, n = 4) or retrograde (RETRO, n = 4) anterior column screw fixation. In a validated setup, a cyclic loading protocol was applied with increasing axial force (750 cycles, 250-750 N) followed by load to failure. Construct survival, energy absorbed, construct stiffness, and load to failure were assessed. Descriptive and opto-metric methods were used to describe the mode of failure. Results All constructs failed with loads below 1500 N. With regard to energy absorbed until failure, the ANTE group resisted to 3.763 x 10(5) N*cycles (range 3.760 x 10(5)-3.763 x 10(5)) and the RETRO group to 3.762 x 10(5) N*cycles (range 3.761 x 10(5)-3.765 x 10(5); p = 1.0). The load to failure was 1254 N (range 977-1299) in the ANTE group and 1234 N (range 1087-1456) in the RETRO group (p = 1.0). Construct stiffness with 250 N was not different between the two groups (ANTE 192 N/mm vs. RETRO 215 N/mm, p = 0.486). In all samples, the mode of failure was a transiliac fracture with screw breakout due to rotation of the pubic fragment around the axis of the screw with a range of rotational motion [ROM] during cyclic testing of 0.96 degrees in one ANTE sample and 0.82 degrees in one RETRO sample for 750 N, and ROM at failure of 2.53 degrees in one ANTE sample and 2.23 degrees in one RETRO sample. There was some plastic deformation of the screws in all cases but no breakage. Conclusions In this in vitro mechanical study, antegrade screw fixation of an anterior column acetabular fracture was not different in construct survival, load to failure, stiffness, and mode of failure when compared to retrograde screw fixation.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据