Repeatability and reproducibility of corneal biomechanical parameters derived from Corvis ST

期刊

EUROPEAN JOURNAL OF OPHTHALMOLOGY
卷 30, 期 6, 页码 1287-1294

出版社

SAGE PUBLICATIONS LTD
DOI: 10.1177/1120672119864554

关键词

Corvis ST; cornea; biomechanics; reproducibility

向作者/读者索取更多资源

Introduction: The aim of this study was to evaluate repeatability and reproducibility of newly calculated biomechanical parameters of the cornea, developed by our research group. Methods: One eye from each of the 23 healthy subjects was measured three times consecutively, three times at different daytimes and on three different days. The within-subject standard deviation and coefficient of variation, as well as the intraclass correlation coefficient, were calculated for every parameter in each group. Results: Excellent repeatability and reproducibility (coefficient of variation < 5%, intraclass correlation coefficient > 0.75) was found for corrected values measured at A1, HC, and A2 time points (2nd A2 Time, 2nd A1 Time, 2nd HC Time, 2nd HC Def Amp and 2nd A1 Def Amp). Corneal-specific stiffness parameters, which showed good repeatability and reliability, were DA_cor (coefficient of variation = 4.02%, intraclass correlation coefficient = 0.919), KcLinear (coefficient of variation = 4.03%, intraclass correlation coefficient = 0.895), areaForceCornea (coefficient of variation = 3.34%, intraclass correlation coefficient = 0.853) and E2 (coefficient of variation = 4.1%, intraclass correlation coefficient = 0.78). Overall, most parameters fell into the category of good reliability (high intraclass correlation coefficient) and poor reproducibility (low coefficient of variation), including all the parameters describing extraocular deformation (DA_ext, AEPvED, AUC EDef, areaForceExtra, Kg and mu g). Comparing the coefficient of variation values for intrasession, intersession and daytime measurements, there were no indices for diurnal changes. Conclusion: Most parameters showed good repeatability and reliability. The extraocular stiffness parameters showed poor reproducibility. KcLinear can serve as a very reliable and repeatable indicator of corneal stiffness.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据