4.4 Article

Design of a Real-time Self-adjusting Calibration Algorithm to Improve the Accuracy of Continuous Blood Glucose Monitoring

期刊

APPLIED BIOCHEMISTRY AND BIOTECHNOLOGY
卷 190, 期 4, 页码 1163-1176

出版社

SPRINGER
DOI: 10.1007/s12010-019-03142-7

关键词

Glucose sensor; Dynamic blood glucose monitoring; Calibration algorithm; Signal drift; Sensitivity attenuation

资金

  1. National Key Research and Development Program of China [2018YFC2000804]
  2. National Natural Science Foundation of China [81671850]
  3. Fundamental Research Funds for the Central Universities [2019CDCG0014]

向作者/读者索取更多资源

The aim of this study is to establish a real-time self-adjusting calibration algorithm to compensate for signal drift and sensitivity attenuation of subcutaneous implantable glucose sensors. A real-time self-adjusting in vivo calibration method was designed based on the one-point calibration model. The current signal was compensated in real-time and the sensitivity was calibrated regularly. The least squares method was used to fit the initial parameters of the model, and then, the in vivo monitored current data was calibrated. Comparing with the mean absolute relative difference (MARD) of the blood glucose concentration by the traditional one-point calibration model (22.85 +/- 5.76%), the MARD of the blood glucose concentration calibrated by the real-time self-adjusting in vivo calibration method was 6.28 +/- 2.31%. The accuracy of the dynamic blood glucose monitoring was effectively improved. This calibration algorithm could compensate the signal drift in real time and correct sensitivity regularly to improve the accuracy of dynamic glucose monitoring, thus significantly enhancing diabetic self-management.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据