4.8 Article

Evaluation of the Performance and Hematocrit Independence of the HemaPEN as a Volumetric Dried Blood Spot Collection Device

期刊

ANALYTICAL CHEMISTRY
卷 91, 期 22, 页码 14467-14475

出版社

AMER CHEMICAL SOC
DOI: 10.1021/acs.analchem.9b03179

关键词

-

资金

  1. Research Foundation-Flanders (FWO) [11F3119N]
  2. Conselleria de Cultura, Educacion e Ordenacion Universitaria, Xunta de Galicia [ED481A-2018/059]

向作者/读者索取更多资源

Dried blood spots (DBS) are often used as a less invasive alternative to venous blood sampling. Despite its numerous advantages, the use of conventional DBS suffers from the hematocrit (hct) effect when analyzing a subpunch. This effect could be avoided by using hct-independent sampling devices, of which the hemaPEN is a recent example. This device collects the blood via four integrated 2.74 mu L microcapillaries, each depositing the blood on a prepunched paper disc. In this study, we evaluated the technical performance of the hemaPEN devices, using an extensive bioanalytical validation and application on authentic patient samples. An LC-MS/MS method quantifying caffeine and its metabolite paraxanthine in dried whole blood (using the hemaPEN device) was fully validated, meeting all preset acceptance criteria. A comparative analysis of 91 authentic patient samples (hct range: 0.17-0.53) of hemaPEN, 3 mm DBS subpunches, and whole blood revealed a limited hct dependence (<= 7% concentration difference over a 0.20-0.50 hct range) for the hemaPEN devices, which we could not attribute to the analytical procedure. Using conventional partial-punch DBS (3 mm punches), concentration differences of >= 25% over this hct range were found. The hemaPEN showed to be robust to the effects of blood sample volume, device lot, analytical operator, and storage stability. The technical performance of the hemaPEN when dealing with patients having a high hct and in cases where a large blood drop is present should be further investigated. Based on the successful validation and application on patient samples, we conclude that the hemaPEN device shows good potential for the volumetric collection of DBS.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据