Liver Stiffness Assessed by Ultrasound Shear Wave Elastography from General Electric Accurately Predicts Clinically Significant Portal Hypertension in Patients with Advanced Chronic Liver Disease

期刊

ULTRASCHALL IN DER MEDIZIN
卷 41, 期 5, 页码 526-533

出版社

GEORG THIEME VERLAG KG
DOI: 10.1055/a-0965-0745

关键词

liver stiffness; ultrasound elastography; portal hypertension; noninvasive

向作者/读者索取更多资源

Purpose Clinically significant portal hypertension (CSPH) is responsible for most of the complications in patients with cirrhosis. Liver stiffness (LS) measurement by vibration-controlled transient elastography (VCTE) is currently used to evaluate CSPH. Bi-dimensional shear wave elastography from General Electric (2D-SWE.GE) has not yet been validated for the diagnosis of PHT. Our aims were to test whether 2D-SWE.GE-LS is able to evaluate CSPH, to determine the reliability criteria of the method and to compare its accuracy with that of VCTE-LS in this clinical setting. Materials and Methods Patients with chronic liver disease referred to hepatic catheterization (HVPG) were consecutively enrolled. HVPG and LS by both VCTE and 2D-SWE. GE were performed on the same day. The diagnostic performance of each LS method was compared against HVPG and between each other. Results 2D-SWE.GE-LS was possible in 123/127 (96.90 %) patients. The ability to record at least 5 LS measurements by 2D-SWE.GE and IQR < 30 % were the only features associated with reliable results. 2D-SWE.GE-LS was highly correlated with HVPG (r = 0.704; p < 0.0001), especially if HVPG < 10mmHg and was significantly higher in patients with CSPH (15.52 vs. 8.14 kPa; p < 0.0001). For a cut-off value of 11.3 kPa, the AUROC of 2D-SWE.GE-LS to detect CSPH was 0.91, which was not inferior to VCTE-LS (0.92; p = 0.79). The diagnostic accuracy of LS by 2D-SWE.GE-LS to detect CSPH was similar with the one of VCTE-LS (83.74 % vs. 85.37 %; p = 0.238). The diagnostic accuracy was not enhanced by using different cut-off values which enhanced the sensitivity or the specificity. However, in the subgroup of compensated patients with alcoholic liver disease, 2D-SWE.GE-LS classified CSPH better than VCTE-LS (93.33 % vs. 85.71 %, p = 0.039). Conclusion 2D-SWE.GE-LS has good accuracy, not inferior to VCTE-LS, for the diagnosis of CSPH.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据