Composition and thickness dependence of TLD relative dose sensitivity: A Monte Carlo study

期刊

RADIATION MEASUREMENTS
卷 129, 期 -, 页码 -

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.radmeas.2019.106191

关键词

Monte Carlo simulation; Dosimeter size; Energy dependence; Beam quality; Perturbation effect

资金

  1. University of Malaya RU Grant [GPF036B-2018]
  2. Malaysia Ministry of Education FRGS Grant [FP032-2017A]

向作者/读者索取更多资源

Absorbed dose sensitivity (S-AD) of a dosimeter results from absorbed dose energy dependence (f) and intrinsic energy dependence (k(bq)) both influenced by compositional and dimensional details. Within attenuating media the SAD is also affected by the depth at which measurements are made as beam quality varies. Monte Carlo simulations allow for studying absorbed dose, importantly providing an effective means by which correction terms for detector response (f(rel)) can be determined. Among the wide range of TLD materials that are available, most generally in particular geometries and size, the commercial products TLD-100 and TLD-100H are perhaps the types most commonly favoured by users. Comprising lithium fluoride (Li natural) co-doped with the activators Mg and Ti or Mg, Cu and P and compressed in a teflon binder to form thicknesses of from 1 mm to a few mm, these have been investigated in previous studies. For some common TLD materials available in small thickness, from 0.125 mm up to 1.000 mm, present study seeks to calculate and compare f(rel) values obtained at different measurement depths in water relative to a reference depth. In simulating a reference radiation field, use has been made of the MCNPX code, calculation providing the correction factors at various depths in a water phantom. Results show that at sub-millimeter dimensions, the effect of detector thickness is more pronounced than the effective atomic number, Z(eff), and density of the TLDs. Smaller thickness and higher Z(eff) TLDs need significant correction factors, with their deviation from unity increasing with the depth up to 15 cm; f(rel) values in the range of 0.96-1.01 were obtained in this study. Perturbation resulting from the presence of the dosimeter was also evaluated in terms of dosimeter size and material composition. Although dosimeter size, density and Z(eff) are well-known parameters, all important in assessing the amount of perturbation, it has been shown herein that the effect of density is more pronounced than Z(eff). For the particular TLDs it is also shown that 0.125 mm thick dosimeters produce a perturbation that might be considered negligible for radiation fields larger than a few centimeters.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据