4.4 Article

Evaluation of a real-time optically stimulated luminescence beryllium oxide (BeO) fibre-coupled dosimetry system with a superficial 140 kVp X-ray beam

期刊

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.ejmp.2019.08.021

关键词

Optically stimulated luminescence; BeO ceramic; Fibre-coupled luminescence dosimetry; Stem effect

资金

  1. University of South Australia [RM39973]

向作者/读者索取更多资源

The purpose of this study was to investigate the potential of real-time optically stimulated luminescence (rtOSL) measurements of a beryllium oxide (BeO) ceramic fibre-coupled luminescence dosimetry system. By pulsing the stimulation laser during the exposure to ionizing radiation, an rtOSL dose-rate measurement can be obtained which could be stem effect free. A portable rtOSL BeO ceramic fibre-coupled dosimetry system is presented and characterized using a constant dose-rate superficial 140 kVp X-ray beam. The rtOSL was measured for dose-rates between 0.29 and 3.88 Gy/min, controlled by varying the source to surface distance. After correcting for OSL decay during the exposure, a linear dose-rate response of the change in rtOSL (Delta rtOSL) was observed. The Delta rtOSL was also observed to be stem effect free.