4.4 Article

High-quality, high-throughput cryo-electron microscopy data collection via beam tilt and astigmatism-free beam-image shift

期刊

JOURNAL OF STRUCTURAL BIOLOGY
卷 208, 期 3, 页码 -

出版社

ACADEMIC PRESS INC ELSEVIER SCIENCE
DOI: 10.1016/j.jsb.2019.09.013

关键词

Cryo-EM data collection; Beam-image shift; Automatic coma determination program; Astigmatism

资金

  1. National Key R&D Program of China [2017YFA0504700]
  2. Strategic Priority Research Program of CAS [XDB08030204]
  3. National Thousand (Young) Talents Program from the Office of Global Experts Recruitment in China

向作者/读者索取更多资源

The throughput of cryo-electron microscopy (cryo-EM) can be improved by employing a procedure that collects beam-image shift data. However, this procedure inadvertently induces a beam tilt, thus decreasing the resolution of the reconstruction. Here, we report an automatic calibration procedure for correcting the beam tilt and a large unexpected astigmatism in the beam-image shift data collection. In this procedure, the changes of the beam tilt and the astigmatism against the beam shift are measured and calibrated. The beam tilt and the astigmatism are corrected by changing the setting of the microscope using predicted values from the calibration. Using our corrected beam-image shift data collection, we found that the resolution remained identical as long as the distance of the beam shift was below 10 pm. The image throughput increases by 80%, with image quality improving by reducing the residual stage drift, thus benefiting the high resolution cryo-EM structure determination. Such a calibration procedure takes about 3 h and can be applied to different microscopes.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据