Mechanical properties and metallurgical features of new and ex vivo used Reciproc Blue and Reciproc

期刊

INTERNATIONAL ENDODONTIC JOURNAL
卷 53, 期 2, 页码 250-264

出版社

WILEY
DOI: 10.1111/iej.13214

关键词

cyclic fatigue; differential scanning calorimetry; focused ion beam analysis; heat treatment; metallographic analysis; usage-induce degradation

资金

  1. FAR project (University of Modena and Reggio Emilia Funds for Research) Microstructural and Mechanical investigation of Nickel-Titanium Endodontic instruments -MMiNTEndo

向作者/读者索取更多资源

Aim To compare the mechanical properties and metallurgic features of new and used Reciproc Blue and Reciproc instruments. Methodology A total of 120 R25 Reciproc Blue and R25 Reciproc instruments were used. The morphological, chemical, mechanical, thermal and phase composition characteristics of new and ex vivo used files were investigated by scanning electron microscopy (FEG-SEM) with energy dispersive X-ray spectroscopy (EDS), focused ion beam analysis (FIB), micro-Raman spectroscopy, FEG-SEM metallography, X-ray diffraction (XRD), differential scanning calorimetry (DSC) and indentation tests. Usage-induce degradation was evaluated. Ten new and ten used instruments per type were run until fracture occurred in a stainless steel artificial canal (60 degrees angle of curvature, 4-mm radius). Time to fracture and the length of the fractured fragment were recorded. Torque and angle of rotation at failure of ten new and ten used instruments for each type were measured according to ISO 3630-1. The fracture surface of each fragment was examined. Two-way analyses of variance was used to analyse the data statistically (alpha-level 0.05). Results SEM analysis revealed microcracks near the tip on both files after ex vivo usage tests. FIB imaging and micro-Raman spectroscopy confirmed the presence of an oxide layer on the Reciproc Blue surface. There was no thinning of the coating after use. XRD revealed a reduction of martensite and R-phase in Reciproc Blue after use. DSC analysis revealed different transformation temperatures for the instruments analysed. Reciproc Blue was significantly more flexible than Reciproc for both new and used samples (P < 0.05), and they were significantly more resistant to cyclic fatigue than Reciproc (P < 0.05). Ex vivo usage reduced the fatigue resistance of both files. Torsional resistance of Reciproc and Reciproc Blue was not reduced by simulated use (P > 0.05). Conclusions The thermal treatment of Reciproc Blue was associated with a finer structure with smaller grains than Reciproc, which increased its fracture resistance and was also responsible for its reduced hardness and lower elastic modulus. Both files were safe during ex vivo usage in severely curved canals.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据