4.4 Article

Energy Calibration of High-Resolution X-Ray TES Microcalorimeters With 3 eV Optical Photons

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TASC.2019.2899856

关键词

Calibration sources; Energy resolution; Microcalorimeters; Transition edge sensors; X-ray detectors

资金

  1. NASA [NNX16AM31G]

向作者/读者索取更多资源

With the improving energy resolution of transition-edge sensor (TES) based microcalorimeters, performance verification and calibration of these detectors have become increasingly challenging, especially in the energy range below 1 keV where fluorescent atomic X-ray lines have linewidths that are wider than the detector energy resolution and require impractically high statistics to determine the gain and deconvolve the instrumental profile. Better behaved calibration sources such as grating monochromators are too cumbersome for space missions and are difficult to use in the lab. As an alternative, we are exploring the use of pulses of 3 eV optical photons delivered by an optical fiber to generate combs of known energies with known arrival times. Here, we discuss initial results of this technique obtained with 2 and 0.7 eV resolution Xray microcalorimeters. With the 2 eV detector, we have achieved photon number resolution for pulses with mean photon number up to 133 (corresponding to 0.4 keV).