4.6 Article

Comparison of Goldmann applanation and Ocular Response Analyser tonometry: intraocular pressure agreement and patient preference

期刊

EYE
卷 34, 期 3, 页码 584-590

出版社

NATURE PUBLISHING GROUP
DOI: 10.1038/s41433-019-0556-2

关键词

-

资金

  1. Atlantic Philanthropies
  2. Economic and Social Research Council
  3. UKCRC Centre of Excellence for Public Health Northern Ireland
  4. Centre for Ageing Research and Development in Ireland
  5. Office of the First Minister and Deputy First Minister
  6. Health and Social Care Research and Development Division of the Public Health Agency
  7. Wellcome Trust/Wolfson Foundation
  8. Queen's University Belfast
  9. Belfast Association
  10. ESRC [ES/L008459/1, ES/S00744X/1] Funding Source: UKRI
  11. MRC [MR/S003770/1] Funding Source: UKRI

向作者/读者索取更多资源

Objectives To evaluate the agreement between Goldmann applanation tonometry (GAT) and Ocular Response Analyser (ORA) intraocular pressure (IOP) measurements, and patients' preferences. Methods Both eyes of participants in the 'Glaucoma within the Northern Ireland Cohort for the Longitudinal Study of Ageing' (GwNICOLA) were included. Participants underwent GAT by a glaucoma expert and ORA tonometry in a random order. Investigators were masked to measurements between devices. Participants were asked which tonometer, if any, they would prefer. We estimated the 95% limits of agreement (95% LoA) and the variables that influence agreement between tonometers. Results There were 228 eyes of 120 participants included in this study. Mean age of participants was 68.0 years (SD 8.79) and 52.5% were female. For GAT-ORA IOPcc the mean difference with GAT (95% CI) was -0.23 mmHg (-0.57 mmHg, 0.11 mmHg) and the 95% LoA (95% CIs) were from 4.82 mmHg (5.15 mmHg, 4.48 mmHg) to -5.28 mmHg (-5.61 mmHg, -4.94 mmHg). 40.8% of eyes had an IOP difference of 2 mmHg or more between GAT and ORA IOPcc. Corneal resistance factor (CRF) as estimated by ORA influenced the agreement between GAT and ORA IOPcc. There were no differences in preference for method of tonometry. Conclusions Although ORA IOPcc measurements with ORA did not show significant bias compared with GAT, the relatively large proportion of measurement differences between ORA IOPcc and GAT that were >2 mmHg indicates that GAT and ORA IOP measurements may not be interchangeable. There were no differences in preference for method of tonometry.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据