Non-Newtonian Blood Flow Simulation of Diastolic Phase in Bileaflet Mechanical Heart Valve Implanted in a Realistic Aortic Root Containing Coronary Arteries

期刊

ARTIFICIAL ORGANS
卷 40, 期 10, 页码 E179-E191

出版社

WILEY
DOI: 10.1111/aor.12787

关键词

Bileaflet mechanical heart valve; Numerical simulation; Non-Newtonian; Coronary arteries; Diastolic phase

向作者/读者索取更多资源

Coronary arteries, which are branched from the sinuses, have tangible effects on the hemodynamic performance of the bileaflet mechanical heart valve (BMHV), especially in the diastolic phase. To better understand this issue, a computer model of ascending aorta including realistic sinus shapes and coronary arteries has been generated in this study in order to investigate the BMHV performance during diastole. Three-dimensional transient numerical analysis is conducted to simulate the diastolic blood flow through the hinges and in coronary arteries under the assumption of non-Newtonian behavior. Results indicate that as blood flows to the coronary arteries mainly during diastole, leakage flow from the hinge and other gaps will change considering the influence of coronary arteries. In addition, BMHV in the case of aortic replacement will increase blood flow rate into the coronary arteries about 100% as the mechanical valve resistance is higher than a native heart valve. Also, it will change the wall shear stress (WSS) distribution and increase coronary artery disease (CAD) potential. It is found out that although less leakage flow reduces the velocity magnitudes through the gaps, the shear stress acting on blood elements with non-Newtonian assumption will be detrimental in the hinge corner at the ventricular side. High WSS of 1800 Pa is observed at beginning of diastole at this region.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据