期刊
JOURNAL OF SYNCHROTRON RADIATION
卷 26, 期 -, 页码 1045-1051出版社
INT UNION CRYSTALLOGRAPHY
DOI: 10.1107/S1600577519003795
关键词
free-electron lasers; photon diagnostics; hard X-rays
X-ray gas monitors (XGMs) are operated at the European XFEL for non-invasive single-shot pulse energy measurements and average beam position monitoring. They are used for tuning and maintaining the self-amplified spontaneous emission (SASE) operation and for sorting single-shot experimental data according to the pulse-resolved energy monitor data. The XGMs were developed at DESY based on the specific requirements for the European XFEL. In total, six XGM units are continuously in operation. Here, the main principle and experimental setup of an XGM are summarized, and the locations of the six XGMs at the facility are shown. Pulse energy measurements at 0.134nm wavelength are presented, exceeding 1mJ obtained with an absolute measurement uncertainty of 7-10%; correlations between different XGMs are shown, from which a SASE1 beamline transmission of 97% is deduced. Additionally, simultaneous position measurements close to the undulator and at the end of the tunnel are shown, along with the correlation of beam position data simultaneously acquired by an XGM and an imager.