4.7 Article

An Unobtrusive and Calibration-free Blood Pressure Estimation Method using Photoplethysmography and Biometrics

期刊

SCIENTIFIC REPORTS
卷 9, 期 -, 页码 -

出版社

NATURE PUBLISHING GROUP
DOI: 10.1038/s41598-019-45175-2

关键词

-

资金

  1. National Key Research and Development Program of China [2018YFC2001100]
  2. Suzhou New District Health Technology Plan [2016Z001]
  3. Suzhou Wuzhong Innovation Talent Award [WC201625]

向作者/读者索取更多资源

We introduce a novel paradigm to unobtrusively and optically measure blood pressure (BP) without calibration. The algorithm combines photoplethysmography (PPG) waveform analysis and biometrics to estimate BP, and was evaluated in subjects with various age, height, weight and BP levels (n = 1249). In the young population (<50 years old) with low, medium and high systolic blood pressures (SBP, <120 mmHg; 120-139 mmHg; >= 140 mmHg), the fitting errors are 6.3 +/- 7.2, -3.9 +/- 7.2 and -20.2 +/- 14.2 mmHg for SBP respectively; In the older population (>50 years old) with the same categories, the fitting errors are 12.8 +/- 9.0, 0.5 +/- 8.2 and -14.6 +/- 11.5 mmHg for SBP respectively. A simple personalized calibration reduces fitting errors significantly (n = 147), and good peripheral perfusion helps to improve the fitting accuracy. In conclusion, PPG may be used to calculate BP without calibration in certain populations. When calibrated, it shows great potential to serially monitor BP fluctuation, which can bring tremendous economic and health benefits.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据