Evaluation of a bilinear model for attenuation correction using CT numbers generated from a parametric method

期刊

APPLIED RADIATION AND ISOTOPES
卷 107, 期 -, 页码 77-86

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.apradiso.2015.09.014

关键词

Attenuation correction; Hybrid systems; Stoichiometric calculation

向作者/读者索取更多资源

A parametric model is used for the calculation of the CT number of some selected human tissues of known compositions (H-i) in two hybrid systems, one SPECT-CT and one PET-CT. Only one well characterized substance, not necessarily tissue-like, needs to be scanned with the protocol of interest. The linear attenuation coefficients of these tissues for some energies of interest (mu(i)) have been calculated from their tabulated compositions and the NIST databases. These coefficients have been compared with those calculated with the bilinear model from the CT number (mu(B)(i)). No relevant differences have been found for bones and lung. In the soft tissue region, the differences can be up to 5%. These discrepancies are attributed to the different chemical composition for the tissues assumed by both methods. (C) 2015 Elsevier Ltd. All rights reserved.