Performance tests of a large volume cerium tribromide (CeBr3) scintillation detector

期刊

APPLIED RADIATION AND ISOTOPES
卷 114, 期 -, 页码 50-56

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.apradiso.2016.04.031

关键词

CeBr3 detector performance tests; Portable neutron generator based; PGNAA setup; CeBr3 detector; Prompt gamma measurements; Boron, cadmium and mercury contaminated water samples; Minimum detection limit of boron, cadmium and mercury in water sample

资金

  1. KACST, Saudi Arabia [ENV2367-04]
  2. Department of Physics, King Fahd University of Petroleum and Minerals, Dhahran, Saudi Arabia
  3. Department of Chemistry, King Fahd University of Petroleum and Minerals, Dhahran, Saudi Arabia

向作者/读者索取更多资源

The response of a large cylindrical 76 mm x 76 mm (height x diameter) cerium tribromide (CeBr3) detector was measured for prompt gamma rays. The total intrinsic activity of the CeBr3 detector, which was measured over 0.33-333 MeV range, was found to be 0.022 +/- 0.001 counts/s/cm(3). The partial intrinsic activity ( due to Ac-227 contamination), was measured over a energy range of 1.22-2.20 MeV energy, was found to be 0.007 +/- 0.001 counts/s/cm(3). Compared to intrinsic activities of LaBr3:Ce and LaCl3:Ce detectors of equivalent volume, the CeBr3 detector has 7-8 times less total intrinsic activity. The detector response for low energy prompt gamma rays was measured over 0.3-0.6 MeV gamma energy range using a portable neutron generator-based Prompt Gamma Neutron Activation Analysis (PGNAA) setup. The experimental yield of boron, cadmium and mercury prompt gamma-rays was measured from water samples contaminated with 0.75-2.5 wt mercury, 031-2.50 wt% boron, and 0.0625-0.500 wt% cadmium, respectively. An excellent agreement has been observed between the calculated and experimental yields of the gamma rays. Also minimum detection limit (MDC) of the CeBr3 detector was measured for boron, cadmium and mercury samples. The CeBr3 detector has 23% smaller value of MDCB and 18% larger value of MDCCd than those of a LaBr3:Ce detector of equivalent size. This study has shown that CeBr3 detector has an excellent response for the low energy prompt gamma-rays with almost an order of magnitude low intrinsic activity as compared to LaCl3:Ce and LaBr3:Ce detectors of equivalent volume. (C) 2016 Elsevier Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据