4.2 Article

Simultaneous Analysis of Iodine and Bromine Species in Infant Formula using HPLC-ICP-MS

期刊

JOURNAL OF AOAC INTERNATIONAL
卷 102, 期 4, 页码 1199-1204

出版社

AOAC INT
DOI: 10.5740/jaoacint.18-0352

关键词

-

向作者/读者索取更多资源

Background: A fast and sensitive method for the simultaneous analysis of iodine and bromine species in infant formula was developed using HPLC-inductively coupled plasma-MS (HPLC-ICP-MS). Method: To determine the four halogen species [iodide (I-), iodate (IO3-), bromide (Br-), and bromate (BrO3-)] in a milk-based Standard Reference Material and four commercially available infant formula products. Results: The four halogen species were baseline separated in less than 6.5 min using an anion exchange resin column and 5.0 mM NaH2PO4 / 15.0 mM Na2SO4 / 5.0 mM EDTA (pH 7.0) mobile phase. Following separation, the halogen species were detected by measuring m/z 79 for Br and m/z 127 for I using a triple quadrupole-ICP-MS. The instrument was operated in single quadrupole mode with helium cell gas. Excellent linearity (R = 0.9999 or better) was obtained for all four species with calibration standards ranging from 0 to 100 ppb. The LOD for I-, IO3-, Br-, and BrO3- were all less than 0.67 mu g/kg. To test the suitability of the method for the accurate determination of low concentrations of the four species in infant formula samples, a spike recovery test was carried out at 20 and 40 mu g/kg into the diluted infant formula samples. Conclusions: Total elemental determinations of iodine and bromine were also performed using the triple quadrupole-ICP-MS without HPLC.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据