4.4 Article

Validation of a Photoplethysmographic Heart Rate Monitor: Polar OH1

期刊

INTERNATIONAL JOURNAL OF SPORTS MEDICINE
卷 40, 期 7, 页码 462-467

出版社

GEORG THIEME VERLAG KG
DOI: 10.1055/a-0875-4033

关键词

photoplethysmography; accuracy; exercise; Bland-Altman; training load; skin type

向作者/读者索取更多资源

This study assessed the validity of a photoplethysmographic heart rate (HR) monitor, the Polar OH1 in various sports performed in ecological conditions: running, cycling, soccer, kayaking, walking, tennis and fitness. Seventy trained athletes (56 males, 14 females) wore the Polar OH1 armband and the H7 chest belt during training. A total of 390h and 38min of recording were compared using a 20-bpm window to assess data quality, and Bland-Altman agreements and ICC analyses were used to test accuracy. Linear regression analyses evaluated the HR accuracy and correlation with skin tone. Training loads (TRIMPs) were compared for each session. Reliability was high for endurance sports (>99%) and lower for sports involving arm movements (92 similar to 95%). Biases were slightly negative for all sports, whereas widths of limits of agreement varied from 7-20bpm. Bland-Altman agreements were all under 5% except tennis, kayak and fitness. HR accuracy was positively correlated to skin tone (p<0.05). Finally, TRIMPs from the OH1 device were inferior to criterion's (except walking and soccer), within a 3% range from reference. Hence, OH1 represents a valid tool to monitor instantaneous HR and training load, especially for endurance sports.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据