4.6 Article

On the Bending and Stretching of Liquid Metal Receive Coils for Magnetic Resonance Imaging

期刊

IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING
卷 66, 期 6, 页码 1542-1548

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TBME.2018.2875436

关键词

Eutectic Gallium-Indium (eGaIn); liquid metal; magnetic resonance imaging (MRI); radio frequency (RF) receive coil; stretchable coil

资金

  1. Nano-Tera Program of the Swiss National Science Foundation

向作者/读者索取更多资源

The eGaIn coil on neoprene demonstrated in this paper presents a stretchable radio frequency receive coil for magnetic resonance imaging (MRI). The coil with dimensions 86 mm x 70 mm is tuned to resonate at 128 MHz for 3 T MRI. We investigate the effect of stretching (up to 40% strain) and bending (50 mm radius of curvature) of the coil on the coil's resistance and resonance frequency. Measurements and simulations show a decrease in resonance frequency of 2.5 MHz per 10% strain. The higher resistivity of liquid metal compared to copper reduces the SNR of MRI scans by 34%; therefore, a tradeoff between flexibility and performance remains. Nevertheless, we have successfully performed MRI scans with the liquid metal coil.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据