Flow rate accuracy of ambulatory elastomeric and electronic infusion pumps when exposed to height and back pressures experienced during home infusion therapy

期刊

EXPERT REVIEW OF MEDICAL DEVICES
卷 16, 期 8, 页码 735-742

出版社

TAYLOR & FRANCIS LTD
DOI: 10.1080/17434440.2019.1632187

关键词

Elastomeric; electronic; infusion device; pump; height; pressure; flow rate

资金

  1. Flinders University Medical Device Partnering Program
  2. South Australian Government, Department of Further Education, Employment, Science and Technology

向作者/读者索取更多资源

Background: Elastomeric infusion pumps are widely used in the delivery of parenteral medications in the home, but real-life conditions may not match calibration or standardised testing conditions. This study investigated the impact of changes in infusion pump height and/or back pressure on infusion pump function. Methods: Volume delivered after one day, infusion duration, average and peak flow rates and time spent within stated accuracy were determined for four elastomeric and one electronic pump using gravimetric technique. Experiments were repeated after altering the height of the pump relative to the output (+/- 40cm, +/- 20cm) and/or adding a back pressure (10-30mmHg) to the output of an attached catheter. Results: Under ideal operating conditions, the flow rate deviated from that specified by the manufacturer and between 88.5% and 99% of the total infusion volume was delivered. Varying the height or applying back pressure led to further changes in average flow rates and the volume of infusion solution delivered by the elastomeric pumps, but had little effect on the electronic pump. Conclusions: Clinicians should consider potential impact on drug delivery, safety and therapeutic effect for home infusion patients given variations in infusion pump performance observed in this study.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据