Improving the detectability of oxygen saturation level targets for preterm neonates: A laboratory test of tremolo and beacon sonifications

期刊

APPLIED ERGONOMICS
卷 56, 期 -, 页码 160-169

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.apergo.2016.03.013

关键词

Patient monitoring; Pulse oximetry; Sonification; Auditory display

资金

  1. Australian Research Council through Discovery Project [DP140101822]

向作者/读者索取更多资源

Recent guidelines recommend oxygen saturation (SpO(2)) levels of 90%-95% for preterm neonates on supplemental oxygen but it is difficult to discern such levels with current pulse oximetry sonifications. We tested (1) whether adding levels of tremolo to a conventional log-linear pulse oximetry sonification would improve identification of SpO(2) ranges, and (2) whether adding a beacon reference tone to conventional pulse oximetry confuses listeners about the direction of change. Participants using the Tremolo (94%) or Beacon (81%) sonifications identified SpO(2) range significantly more accurately than participants using the LogLinear sonification (52%). The Beaton sonification did not confuse participants about direction of change. The Tremolo sonification may have advantages over the Beacon sonification for monitoring SpO(2) of preterm neonates, but both must be further tested with clinicians in clinically representative scenarios, and with different levels of ambient noise and distractions. Crown Copyright (C) 2016 Published by Elsevier Ltd. All rights reserved.