4.3 Article

Validity of single-channel masseteric electromyography by using an ultraminiature wearable electromyographic device for diagnosis of sleep bruxism

期刊

JOURNAL OF PROSTHODONTIC RESEARCH
卷 64, 期 1, 页码 90-97

出版社

ELSEVIER IRELAND LTD
DOI: 10.1016/j.jpor.2019.04.003

关键词

Sleep bruxism; Polysomnography; Single-channel electromyogram; Sensitivity and specificity; Cut-off value

资金

  1. Hokkaido University Hospital
  2. Graduate School of Dental Medicine of Hokkaido University

向作者/读者索取更多资源

Purpose: The purpose of this study was to clarify the validity of assessment of sleep bruxism (SB) by using single-channel electromyogram (EMG) and a cut-off value with optimum sensitivity and specificity. Methods: The subjects were twenty volunteers with clinical diagnosis of SB. Assessment by masseteric EMG data only by using a wearable EMG was compared with reference standard assessment by masseteric EMG data using polysomnography with audio-visual recording (PSG-AV). From EMG activities recorded by single-channel EMG, bursts of more than two times the baseline amplitude with a duration of 0.25 s or more were selected by a burst unit (EMG-burst-all). Furthermore, from EMG-burst-all, bursts that were more than 5-20% of the maximum voluntary contraction value (EMG-burst-5%, EMG-burst-10%, EMG-burst-20%) were selected. By an episode unit, phasic, tonic, and mixed episodes were selected by single-channel EMG (EMG-episodes). Among the EMG-episodes, further, reference standard episodes of SB (PSG-episodes) were selected by PSG-AV assessment. Results: Sixteen subjects were diagnosed as bruxers based on PSG-AV (PSG-episodes/h >2). By a burst unit and an episode unit, there were significant correlations between assessment variables of SB by single-channel EMG and PSG-AV except for EMG-burst-20%/h. When the cut-off value in EMG-episodes/h was 5.5/h for sleep bruxers, both sensitivity and specificity were 100%. As for variables by a burst unit, EMG-burst-all/h and EMG-burst-5% had higher values of sensitivity and specificity. Conclusions: The results suggested that single-channel EMG is valid for diagnosis of SB if a cut-off value that is appropriate for single-channel EMG is used. (C) 2019 Japan Prosthodontic Society. Published by Elsevier Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据