Quantification below the LLOQ in regulated LC-MS/MS assays: a review of bioanalytical considerations and cautions

期刊

BIOANALYSIS
卷 11, 期 8, 页码 797-814

出版社

FUTURE SCI LTD
DOI: 10.4155/bio-2018-0261

关键词

below LLOQ; BLQ estimations; BLQ extrapolations; BLQ values; PK modeling; regulated bioanalysis

向作者/读者索取更多资源

In response to an earlier workshop covering the pros and cons of quantification below the LLOQ (BLQ) the author reviews the topics discussed from the bioanalytical standpoint. Important considerations for estimating concentrations below the LLOQ include: method signal-to-noise, baseline shape and condition, close lying interference peaks (especially for protein methods), matrix effect, adsorption and stability of the analyte at low concentrations and carryover. These methodological issues are discussed as possible contributors to inaccuracy in BLQ estimations, and appropriate cautions are provided via examples. A proposed method for the evaluation of BLQ estimations utilizes extended incurred sample reanalysis analysis where BLQ samples or spiked simulated samples are analyzed with quality controls and standards in addition to those in the original study. Generally, BLQ estimations are discouraged, with the recommendation that any extrapolations should be done in close collaboration between the pharmacokinetic (PK) and bioanalytical scientists in consultation with the regulatory agency.