Detection of 1 zmol injection of angiotensin using capillary zone electrophoresis coupled to a Q-Exactive HF mass spectrometer with an electrokinetically pumped sheath-flow electrospray interface

期刊

TALANTA
卷 204, 期 -, 页码 70-73

出版社

ELSEVIER
DOI: 10.1016/j.talanta.2019.05.079

关键词

Electrospray ionization; Capillary zone electrophoresis; Zeptomole detection; Sub-zeptomole detection limit

资金

  1. National Institutes of Health [R01GM096767]

向作者/读者索取更多资源

An electrokinetically pumped sheath-flow nanoelectrospray interface provides an efficient means of transferring ions from a capillary electrophoresis system into a mass spectrometer. To characterize its performance, we analyzed angiotensin solutions prepared in a background of 0.25 mg/mL of a BSA tryptic digest. Calibration curves were prepared from 10 zmol (10(-2)1 mol) to 10 fmol (10(-1)4 mol) of angiotensin injected into the capillary. A parallel reaction monitoring approach was used; MS1 was set to m/z = 523.8 +/- 2, and fragment ion signals at 263.1389 (y(2+)) and 784.4095 (b(6+)) were used to generate selected ion electropherograms. Calibration curves based on peak areas were linear (log-log slope of 0.94 for the y(2+) fragment and 0.98 for the b(6+) fragment). We then injected 1-zmol (600 copies) of angiotensin in the BSA background using a 10- mu m ID separation capillary. Triplicate analyses consistently produced co-migrating peaks for the two fragment ions. The b(6+) electropherogram showed no background signal, whereas the y(2+) electropherogram showed a few noise spikes that were smaller than the peak maximum. Bienayme-Tchebycheff inequality was used to estimate detection limits of 230 ymol (140 ions) injected into the separation capillary. To the best of our knowledge, these electropherograms present the smallest number of molecules detected using mass spectrometry coupled with a separation.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据