Trends in radiation dose and image quality for pediatric patients with a multidetector CT and a third-generation dual-source dual-energy CT

期刊

RADIOLOGIA MEDICA
卷 124, 期 8, 页码 745-752

出版社

SPRINGER-VERLAG ITALIA SRL
DOI: 10.1007/s11547-019-01037-5

关键词

Dual-source CT; Dual-energy CT; Pediatric radiology; Dose; Dose reduction

向作者/读者索取更多资源

Aim To provide an overview on dose reduction and image quality after the installation of a third-generation dual-source CT (dsCT) in a Pediatric Radiology Department. Materials and methods We included pediatric patients (< 20 years old) undergoing CT for oncological staging (neck, chest and abdomen) or low-dose chest CT for lung diseases. Each of these two groups were further divided in two age groups (<= or > 10 years old) including patients scanned in the same period of two consecutive years, in 2017 with a 16-row LightSpeed CT (GE Healthcare) or in 2018 with a Somatom Force dsCT (Siemens Healthineers). Technical parameters such as kVp, mAs, slice thickness, exposure times and dose indicators were retrieved and compared. Image quality was evaluated in consensus by two radiologists on a five-point semiquantitative scale. Nonparametric tests were used. Results In oncological patients, significantly lower kVp and tube current with better image quality were achieved with the dsCT. Radiation dose (total DLP) was 5-6 times lower with dsCT, thanks also to virtual non-contrast images. In low-dose chest CT, the frequent use of tin filter required higher tube current; a total DLP 3 times lower was achieved with dsCT in patients <= 10 years old. The image quality was better with the dsCT in low-dose chest CT protocols. Conclusion The third-generation dsCT provides high-quality images with reduced motion artifacts at lower dose.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据