JOURNAL OF CHROMATOGRAPHY A
卷 1603, 期 -, 页码 412-416出版社
ELSEVIER
DOI: 10.1016/j.chroma.2019.05.040
关键词
Column packing; Flow-reversal; Packed bed heterogeneity; Sub-2-mu m particles; UHPLC
资金
- NKFIH OTKA [K125312]
- New National Excellence Program of the Ministry of Human Capacities [UNKP-18-3-I-PTE-125]
The effect of column end structure and bed heterogeneity of six commercially available reversed-phase chromatographic columns for fast liquid chromatography with different column packing materials - such as fully porous (Waters XBridge C-18 with 1.7 mu m particles) and superficially porous (Waters CORTECS C-18 with 1.6 mu m particles), with column dimension of 2.1 x 50,100 or 150 mm were tested with flow-reversal method. The method includes arresting the flow when a non-retained marker (thiourea) has penetrated to a given distance into the column and then reversing the column. Hence, when the flow has been restarted, the sample is eluted at the same end of the column where it entered. The experiments showed that all columns are axially heterogeneous, and some differences could be observed between the two respective column ends. Furthermore, we can conclude that the shorter columns are axially more homogeneous than the longer ones, thus the column length is an influencing factor on the column packing procedure. (C) 2019 Elsevier B.V. All rights reserved.