4.6 Article

The effect of column packing procedure on column end efficiency and on bed heterogeneity - Experiments with flow-reversal

期刊

JOURNAL OF CHROMATOGRAPHY A
卷 1603, 期 -, 页码 412-416

出版社

ELSEVIER
DOI: 10.1016/j.chroma.2019.05.040

关键词

Column packing; Flow-reversal; Packed bed heterogeneity; Sub-2-mu m particles; UHPLC

资金

  1. NKFIH OTKA [K125312]
  2. New National Excellence Program of the Ministry of Human Capacities [UNKP-18-3-I-PTE-125]

向作者/读者索取更多资源

The effect of column end structure and bed heterogeneity of six commercially available reversed-phase chromatographic columns for fast liquid chromatography with different column packing materials - such as fully porous (Waters XBridge C-18 with 1.7 mu m particles) and superficially porous (Waters CORTECS C-18 with 1.6 mu m particles), with column dimension of 2.1 x 50,100 or 150 mm were tested with flow-reversal method. The method includes arresting the flow when a non-retained marker (thiourea) has penetrated to a given distance into the column and then reversing the column. Hence, when the flow has been restarted, the sample is eluted at the same end of the column where it entered. The experiments showed that all columns are axially heterogeneous, and some differences could be observed between the two respective column ends. Furthermore, we can conclude that the shorter columns are axially more homogeneous than the longer ones, thus the column length is an influencing factor on the column packing procedure. (C) 2019 Elsevier B.V. All rights reserved.