期刊
GEOSTANDARDS AND GEOANALYTICAL RESEARCH
卷 43, 期 3, 页码 385-395出版社
WILEY
DOI: 10.1111/ggr.12279
关键词
biomineral; atom probe tomography; sample preparation; electron microscopy; focused ion beam
In this study atom probe tomography was used to study nacre, an important biocomposite material that is challenging to prepare and analyse by atom probe and, when successful, yields data that is challenging to interpret. It was found that these challenges mainly arise from the insulating and heterogeneous nano-scale properties of nacre. We outline our current best practice for preparing and running atom probe tips, such as using a low acceleration voltage (< 3 kV) and current (<= 50 pA) to avoid damage to the microstructure, and using transmission electron microscopy to confirm that the region of interest is located close to the apex of the atom probe tip. Optimisation of the preparation parameters led to several successful atom probe experiments, with one of the data sets containing part of an organic membrane and others showing organic inclusions within the reconstruction.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。