4.6 Article

A simple realtime algorithm for automatic external defibrillator

期刊

BIOMEDICAL SIGNAL PROCESSING AND CONTROL
卷 51, 期 -, 页码 277-284

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.bspc.2019.02.030

关键词

Automatic external defibrillators; Ventricular fibrillation; Ventricular tachycardia; Adaptive threshold; Support vector machine; Hierarchical; Real time; Python; Embedded system; Raspberry pi

向作者/读者索取更多资源

Automatic External Defibrillator is a device that automatically diagnoses the heart rhythm which requires an electric shock. In general the signal processing unit of an AED aims to immediately identify the occurrence of ventricular fibrillation and ventricular tachycardia in a patient's electrocardiograph signal. In this research, a quick and high-precision method is presented that identifies peaks and heart rate arrhythmias of the cardiac signals that do not include QRS complex. Initially, asystole and presence of extra noise are recognized in a preprocessing step. Furthermore, two support vector machine classifiers have been used in a hierarchical way to separate Ventricular Fibrillation, Ventricular Tachycardia and normal signals. As a result, a real-time algorithm is developed that matches American Heart Association standard with an accuracy of 98.6%. Finally the Raspberry pi board is used as a hardware platform to embed the proposed algorithm into an AED system. The results show that the implemented system can detect shock-required rhythms in 0.56 s; and the method may be used in an actual signal processing unit of an AED. (C) 2019 Elsevier Ltd. All rights reserved.