期刊
ANALYTICAL CHEMISTRY
卷 91, 期 11, 页码 7336-7345出版社
AMER CHEMICAL SOC
DOI: 10.1021/acs.analchem.9b01100
关键词
-
资金
- NIST Biomanfacturing Program
- National Institute of General Medical Sciences of the National Institutes of Health (NIH) [U54 GM094586]
- Singapore Ministry of Education Academic Research Fund Tier 3 [MOE2012-T3-1-008]
- Danish Council for Independent Research (Sapere Aude Grant) [DFF-4184-00537A]
- Univ. of Maryland Baltimore, School of Pharmacy Mass Spectrometry Center [SOP1841-IQB2014]
- National Institutes of Health [AI068730, AI030040]
Hydrogen-deuterium exchange mass spectrometry (HDX-MS) is an established, powerful tool for investigating protein-ligand interactions, protein folding, and protein dynamics. However, HDX-MS is still an emergent tool for quality control of biopharmaceuticals and for establishing dynamic similarity between a biosimilar and an innovator therapeutic. Because industry will conduct quality control and similarity measurements over a product lifetime and in multiple locations, an understanding of HDX-MS reproducibility is critical. To determine the reproducibility of continuous-labeling, bottom-up HDX-MS measurements, the present interlaboratory comparison project evaluated deuterium uptake data from the Fab fragment of NISTmAb reference material (PDB: 5K8A) from 15 laboratories. Laboratories reported similar to 89 800 centroid measurements for 430 proteolytic peptide sequences of the Fab fragment (similar to 78 900 centroids), giving similar to 100% coverage, and similar to 10 900 centroid measurements for 77 peptide sequences of the Fc fragment. Nearly half of peptide sequences are unique to the reporting laboratory, and only two sequences are reported by all laboratories. The majority of the laboratories (87%) exhibited centroid mass laboratory repeatability precisions of < s(Lab)> <= (0.15 +/- 0.01) Da (1 sigma((x) over bar)). All laboratories achieved < s(Lab)> <= 0.4 Da. For immersions of protein at T-HDx = (3.6 to 25) degrees C and for D2O exchange times of t(HDX) = (30 s to 4 h) the reproducibility of back-exchange corrected, deuterium uptake measurements for the 15 laboratories is sigma(Laboratories)(reproducibility15) (t(HDX)) = (9.0 +/- 0.9) % (1 sigma). A nine laboratory cohort that immersed samples at T-HDX = 25 degrees C exhibited reproducibility of sigma(25C cohort)(reproducibility) (t(HDX)) = (6.5 +/- 0.6) % for back-exchange corrected, deuterium uptake measurements.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据