AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS
卷 155, 期 4, 页码 592-599出版社
MOSBY-ELSEVIER
DOI: 10.1016/j.ajodo.2018.09.011
关键词
-
资金
- Virginia Commonwealth University Alexander Research Fund
- Southern Association of Orthodontists Research Fund
Introduction: In the literature, there is little information available on 3D-printed orthodontic retainers. This study examined the accuracy of 3D-printed retainers compared with conventional vacuum-formed and commercially available vacuum-formed retainers. Methods: Three reference models (models 1, 2, and 3) were used to fabricate traditional vacuum-formed, commercially available vacuum-formed, and 3D-printed retainers. For each model, retainers were made using the 3 methods (a total of 27 retainers). To determine the trueness, ie, closeness of a model to a true model, the distance between the retainer and its digital model at reference points were calculated with the use of engineering software. The measurements were reported as average absolute observed values and compared with those of the conventional vacuum-formed retainers. Results: Average differences of the conventional vacuum-formed retainers ranged from 0.10 to 0.20 mm. The commercially available and 3D-printed retainers had ranges of 0.10 to 0.30 mm and 0.10 to 0.40 mm, respectively. Conclusions: The conventional vacuum-formed retainers showed the least amount of deviation from the original reference models and the 3D-printed retainers showed the greatest deviation. However, all 3 methods yielded measurements within 0.5 mm, which has previously been accepted to be clinically sufficient.