4.6 Article

A resistance-sensing mechanical injector for the precise delivery of liquids to target tissue

期刊

NATURE BIOMEDICAL ENGINEERING
卷 3, 期 8, 页码 621-631

出版社

NATURE PUBLISHING GROUP
DOI: 10.1038/s41551-019-0350-2

关键词

-

资金

  1. Boston-KPro research fund
  2. [R01HL095722]

向作者/读者索取更多资源

The precision of the delivery of therapeutics to the desired injection site by syringes and hollow needles typically depends on the operator. Here, we introduce a highly sensitive, completely mechanical and cost-effective injector for targeting tissue reliably and precisely. As the operator pushes the syringe plunger, the injector senses the loss-of-resistance on encountering a softer tissue or a cavity, stops advancing the needle and delivers the payload. We demonstrate that the injector can reliably deliver liquids to the suprachoroidal space-a challenging injection site that provides access to the back of the eye-for a wide range of eye sizes, scleral thicknesses and intraocular pressures, and target sites relevant for epidural injections, subcutaneous injections and intraperitoneal access. The design of this simple and effective injector can be adapted for a broad variety of clinical applications.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据