期刊
ACS SENSORS
卷 4, 期 2, 页码 326-334出版社
AMER CHEMICAL SOC
DOI: 10.1021/acssensors.8b00942
关键词
neutravidin biosensor; multiplex detection; microRNAs; molecular beacons; nanoparticle labels
资金
- EU-FP7 project [PIRSES-GA-2012-318053]
Current electrochemical biosensors for multiple miRNAs require tedious immobilization of various nucleic acid probes. Here, we demonstrate an innovative approach using a generic neutravidin biosensor combined with electrochemically encoded responsive nanolabels for facile and simultaneous multiplexed detection of miRNA-21 and miRNA-141. The selectivity of the biosensor arises from the intrinsic properties of the electrochemically encoded responsive nanolabels, comprising biotinylated molecular beacons (biotin-MB) and metal nanoparticles (metal-NPs). The procedure is a simple one-pot assay, where the targeted miRNA causes the opening of biotin-MB followed by capturing of the biotin-MB-metal-NPs by the neutravidin biosensor and simultaneous detection of the captured metal-NPs by stripping square-wave voltammetry (SSWV). The multiplexed detection of miRNA-21 and miRNA-141 is achieved by differentiation of the electrochemical signature (i.e., the peak current) for the different metal-NP labels. The biosensor delivers simultaneous detection of miRNAs with a linear range of 0.5-1000 pM for miRNA-21 and a limit of detection of 0.3 pM (3 sigma/sensitivity, n = 3), and a range of 50-1000 pM for miRNA-141, with a limit of detection of 10 pM. Furthermore, we demonstrate multiplexed detection of miRNA-21 and miRNA-141 in a spiked serum sample.