4.6 Article

A Novel Device for the Evaluation of Hemostatic Function in Critical Care Settings

期刊

ANESTHESIA AND ANALGESIA
卷 123, 期 6, 页码 1372-1379

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1213/ANE.0000000000001413

关键词

-

资金

  1. National Heart, Lung, and Blood Institute of the NIH [R44HL103030]

向作者/读者索取更多资源

Major surgical procedures often result in significant intra-and postoperative bleeding. The ability to identify the cause of the bleeding has the potential to reduce the transfusion of blood products and improve patient care. We present a novel device, the Quantra Hemostasis Analyzer, which has been designed for automated, rapid, near-patient monitoring of hemostasis. The Quantra is based on Sonic Estimation of Elasticity via Resonance Sonorheometry, a proprietary technology that uses ultrasound to measure clot time and clot stiffness from changes in viscoelastic properties of whole blood during coagulation. We present results of internal validation and analytical performance testing of the technology and demonstrate the ability to characterize the key functional components of hemostasis.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据