4.6 Article

Off-beam QEPAS sensor using an 11-μm DFB-QCL with an optimized acoustic resonator

期刊

OPTICS EXPRESS
卷 27, 期 5, 页码 7435-7446

出版社

OPTICAL SOC AMER
DOI: 10.1364/OE.27.007435

关键词

-

类别

资金

  1. Agence Nationale de la Recherche (MULTIPAS project) [ANR-16-CE04-0012]
  2. French Investment for the Future program [EquipEx EXTRA] [ANR-11-EQPX-0016]

向作者/读者索取更多资源

An off-beam quartz-enhanced photoacoustic spectroscopy (QEPAS) sensor was designed for ethylene detection using a distributed-feedback quantum cascade laser (QCL) operating in the mid-infrared around 11 mu m. The acoustic microresonator configuration was experimentally optimized using an original open-cell photoacoustic setup with a MEMS microphone. Correction factors based on theoretical acoustic models were introduced in order to accurately describe the response of millimeter-sized acoustic resonators. The optimized QEPAS sensor exhibited a limit of detection of 60 ppb for 60 s integration, giving a NNEA of 4.8 x 10(-8) W.cm(-1).Hz(-0.5). (C) 2019 Optical Society of America under the terms of the OSA Open Access Publishing Agreement

作者

我是这篇论文的作者