A novel approach to Lab-In-Syringe Head-Space Single-Drop Microextraction and on-drop sensing of ammonia

期刊

ANALYTICA CHIMICA ACTA
卷 934, 期 -, 页码 132-144

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.aca.2016.06.039

关键词

Lab-In-Syringe; Head-space Single-Drop Microextraction; Ammonia; Kinetic on-drop sensing; Bromothymol blue indicator; Automation of sample preparation

资金

  1. Australian government [4713_2015]
  2. Grant Agency of the Czech Republic (GACR) [S15-10781S]

向作者/读者索取更多资源

A novel approach to the automation technique Lab-In-Syringe, also known as In-Syringe Analysis, is proposed which utilizes a secondary inlet into the syringe void, used as a size-adaptable reaction chamber, via a channel passing through the syringe piston. This innovative approach allows straightforward automation of head-space single-drop microextraction, involving accurately controlled drop formation and handling, and the possibility of on-drop analyte quantification. The syringe was used in upside-down orientation and in-syringe magnetic stirring was carried out, which allowed homogenous mixing of solutions, promotion of head-space analyte enrichment, and efficient syringe cleaning. The superior performance of the newly developed system was illustrated with the development of a sensitive method for total ammonia determination in surface waters. It is based on head-space extraction of ammonia into a single drop of bromothymol blue indicator created inside the syringe at the orifice of the syringe piston channel and on-drop sensing of the color change via fiber optics. The slope of the linear relationship between absorbance and time was used as the analytical signal. Drop formation and performance of on-drop monitoring was further studied with rhodamine B solution to give a better understanding of the system's performance. A repeatability of 6% RSD at 10 mmol L-1 NH3, a linear range of up to 25 mmol L-1 NH3, and a limit of detection of 1.8 mmol L-1 NH3 were achieved. Study of interferences proved the high robustness of the method towards humic acids, high sample salinity, and the presence of detergents, thus demonstrating the method superiority compared to the state-of-the-art gas-diffusion methods. A mean analyte recovery of 101.8% was found in analyzing spiked environmental water samples. (C) 2016 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据