pacemaker; cardiovascular implantable electronic device; implantable cardioverter defibrillator; anesthesia; device programming; Biotronik; CLS; perioperative management
Biotronik cardiovascular implantable electronic devices, specifically Biotronik pacemakers, contain unique features that are relevant to perioperative management. For example, Biotronik pacemakers have a programmable response to magnet application, a default magnet response that does not result in sustained asynchronous pacing, and a unique method of rate adaptation (eg, closed loop stimulation). This review article focuses on these unique features; the interpretation of Biotronik interrogation reports; and the basic programming (eg, mode, rate, rate adaptation, tachyarrhythmia therapies) relevant to the perioperative management of Biotronik cardiovascular implantable electronic devices. Published by Elsevier Inc.