Transcatheter Aortic Valve Replacement With Next-Generation Self-Expanding Devices A Multicenter, Retrospective, Propensity-Matched Comparison of Evolut PRO Versus Acurate neo Transcatheter Heart Valves

期刊

JACC-CARDIOVASCULAR INTERVENTIONS
卷 12, 期 5, 页码 433-443

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.jcin.2018.11.036

关键词

Acurate neo; Evolut PRO; self-expanding; TAVR; VARC-2

向作者/读者索取更多资源

OBJECTIVES The aim of this study was to compare transcatheter aortic valve replacement (TAVR) with the Acurate neo (NEO) and Evolut PRO (PRO) devices. BACKGROUND The NEO and PRO bioprostheses are 2 next-generation self-expanding devices developed for TAVR. METHODS The NEOPRO (A Multicenter Comparison of Acurate NEO Versus Evolut PRO Transcatheter Heart Valves) registry retrospectively included patients who underwent transfemoral TAVR with either NEO or PRO valves at 24 centers between January 2012 and March 2018. One-to-one propensity score matching resulted in 251 pairs. Pre-discharge and 30-day Valve Academic Research Consortium (VARC)-2 defined outcomes were evaluated. Binary logistic regression was performed to adjust the treatment effect for propensity score quintiles. RESULTS A total of 1,551 patients (n = 1,263 NEO; n = 288 PRO) were included. The mean age was 82 years, and the mean Society of Thoracic Surgeons score was 5.1%. After propensity score matching (n = 502), VARC-2 device success (90.6% vs. 91.6%; p = 0.751) and pre-discharge moderate to severe (IIthorn) paravalvular aortic regurgitation (7.3% vs. 5.7%; p = 0.584) were comparable between the NEO and PRO groups. Furthermore, there were no significant differences in any 30-day clinical outcome between matched NEO and PRO pairs, including all-cause mortality (3.2% vs. 1.2%; p = 0.221), stroke (2.4% vs. 2.8%; p = 1.000), new permanent pacemaker implantation (11.0% vs. 12.8%; p = 0.565), and VARC-2 early safety endpoint (10.6% vs. 10.4%; p = 1.000). Logistic regression on the unmatched cohort confirmed a similar risk of VARC-2 device success, paravalvular aortic regurgitation IIthorn, and 30-day clinical outcomes after NEO and PRO implantation. CONCLUSIONS In this multicenter registry, transfemoral TAVR with the NEO and PRO bioprostheses was associated with high device success, acceptable rates of paravalvular aortic regurgitation IIthorn, and good 30-day clinical outcomes. After adjusting for potential confounders, short-term outcomes were similar between the devices. (c) 2019 Published by Elsevier on behalf of the American College of Cardiology Foundation.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据