Spinal phantom comparability study of Cobb angle measurement of scoliosis using digital radiographic imaging

期刊

JOURNAL OF ORTHOPAEDIC TRANSLATION
卷 15, 期 -, 页码 81-90

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.jot.2018.09.005

关键词

Cobb angle; Comparison study; Imaging evaluation; Scoliosis; Spine phantom

资金

  1. Department of Diagnostic Radiology, The Duchess of Kent Children's Hospital at Sandy Bay
  2. Department of Health Technology and Informatics, The Hong Kong Polytechnic University

向作者/读者索取更多资源

Background: Computed radiography (CR), digital radiography (DR) and biplanar radiography (EOS (TM) imaging system) are common imaging tools for radiographic evaluation of adolescent idiopathic scoliosis (AIS). The effect of imaging methods in relation to later-on Cobb angle measurements on radiographs is not yet quantified. The study aimed to examine the compatibility between CR, DR and EOS for scoliotic quantification by evaluating the reliability, agreement of different imaging methods, and assessing the prediction performance for EOS measurement from that of CR and DR. Method: A flexible spine phantom was used to simulate 32 different scoliotic curves ranging from 10 degrees to 60 degrees. Each curvature was imaged using DR, CR and EOS systems accordingly. Each of the six observers independently measured Cobb angle twice on each image at a two-week interval. Intraclass correlation coefficient (model 2 and 3), Bland-Altman plot and linear regression analysis were completed to evaluate the reliability, agreement, and the prediction of Cobb angle measurement, respectively. Results: Reliability analysis showed excellent intra-observer reliability (Intraclass correlation coefficient >0.9) for each observer and good inter-observer reliability (Intraclass correlation coefficient = 0.84 for EOS; 0.739 for CR; 0.877 for DR) for each method. Bland-Altman plots demonstrated good agreement between imaging methods without fixed or proportional bias. Excellent coefficient of determination was achieved, with 0.980 for CR versus EOS measurements, and 0.973 for DR versus EOS measurements. Conclusions: Radiographs produced by all of the three methods can provide reliable and accurate Cobb angle measurements for scoliosis assessments. None of the methods systemically underestimates or overestimates the Cobb angle measurement. Additionally, all of the evaluated methods are satisfactory in obtaining images for Cobb angle measurement in AIS. However, the 3D post-processing techniques offered by EOS should also be taken into consideration as it takes a vital role in treatment and monitoring of 3D deformity in the case of scoliosis. The translational potential of this article: In view of the limited availability of biplanar radiography (EOS (TM) imaging system), computed radiography and digital radiography are demonstrated to be reliable alternatives in scoliosis monitoring as evident in the reliability, agreement and prediction of Cobb angle measurement. (C) 2018 The Authors. Published by Elsevier (Singapore) Pte Ltd on behalf of Chinese Speaking Orthopaedic Society.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据