4.6 Article

Mechanical, Structural, and Chemical Analysis of Athrombogenic Multilayer Wall of Ventricular Assist Device Tested in Hydrodynamic Fatigue Tests

期刊

ADVANCED ENGINEERING MATERIALS
卷 18, 期 5, 页码 795-802

出版社

WILEY-V C H VERLAG GMBH
DOI: 10.1002/adem.201500492

关键词

-

资金

  1. National Centre for Research and Development in Poland [PBS1/A3/11/2012]
  2. Ministry of Science and Higher Education in Poland [11.11.110.295]

向作者/读者索取更多资源

The purpose of the paper is to assess an impact of short- and long-term operational load on continuity and composition of TiN/Au layers deposited on the polymer Bionate II that the pulsatile ventricular assist device Religa Heart Ext is made of. The additional objective is to determine mechanical properties of the TiN/Au layers deposited on the polymer. The physical model of circulatory system was built using the Windkessel model. The tests with the use of energy dispersive spectroscopy and scanning electron microscopy were made on surfaces of samples made of the Religa Heart Ext material examined in the fatigue tests. Mechanical properties of the TiN/Au coatings deposited on the polymer were measured in the nanoindentation test by comparison of the coated and the uncoated samples.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。