Evaluation of the Seegene Allplex? Respiratory Panel for diagnosis of acute respiratory tract infections

期刊

ACTA CLINICA BELGICA
卷 74, 期 6, 页码 379-385

出版社

TAYLOR & FRANCIS LTD
DOI: 10.1080/17843286.2018.1531605

关键词

Respiratory pathogens; molecular detection; multiplex PCR; real-time PCR

向作者/读者索取更多资源

Objectives: The Seegene Allplex Respiratory panel was retrospectively challenged using a collection of quality control samples (QCMD) and clinical samples previously analysed with validated routine methods. Methods: A collection of 111 samples [43 QCMD samples, 13 bronchoalveolar lavage fluids and 55 nasopharyngeal aspirates/swabs] was tested with Seegene Allplex. The clinical samples were tested previously using either FTD? Respiratory Pathogens 21 qPCR assay (Fast Track Diagnostics), an in-house multiplex PCR for Bordetella, or BioGX Sample-Ready(TM) Atypical pneumo panel (Becton Dickinson). Samples were stored at ?80?C prior to analysis with Seegene Allplex?, nucleic acids were automatically extracted with NucliSENS Easymag (bioM?rieux). Samples returning discordant results were subjected to repeat testing and/or additional testing by reference laboratories. Results: Seegene correctly identified 41/43 QCMD samples (95.4%); two samples positive for respiratory syncytial virus (RSV) and human metapneumovirus, respectively, were only correctly identified following repeat testing. In the 56 clinical samples, overall, 97 pathogens were identified: 65 pathogens (67.0%) were detected both by routine methods and Seegene, 24 pathogens (24.7%) only by routine methods, and 8 pathogens (8.2%) only by Seegene. The majority of discordant results was detected in samples with low pathogen load (22/32, 68.8%) and in samples containing multiple pathogens (25/32, 78.1%). Full agreement between methods was observed for influenza, RSV, adenovirus, Bordetella (para)pertussis and Chlamydia pneumoniae. Discordance was observed for human metapneumovirus, coronavirus OC43,?bocavirus?and parainfluenza virus, mainly type 4. Conclusion: Overall, the Seegene Allplex(TM) assay performed well for routine detection of important respiratory targets. Acceptable agreement was observed between Seegene and other routine assays.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.0
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据