4.5 Article Proceedings Paper

High resolution, high signal-to-noise crystal spectrometer for measurements of line shifts in high-density plasmas

期刊

REVIEW OF SCIENTIFIC INSTRUMENTS
卷 89, 期 10, 页码 -

出版社

AMER INST PHYSICS
DOI: 10.1063/1.5035303

关键词

-

资金

  1. U.S. DOE by LLNL [DE-AC52-07NA27344]

向作者/读者索取更多资源

The Orion high-resolution x-ray (OHREX) spectrometer has been a successful tool for measuring the shapes of density-broadened spectral lines produced in short-pulse heated plasmas at the Orion laser facility. We have recently outfitted the instrument with a charge-couple device (CCD) camera, which greatly increased the accuracy with which we can perform line-shift measurements. Because OHREX is located on the outside of the Orion target chamber, no provisions for the shielding of electromagnetic pulses are required. With the CCD, we obtained a higher signal-to-noise ratio than we previously obtained with an image-plate detector. This allowed us to observe structure in the image produced by the diffraction from the two OHREX crystals, which was highly reproducible from shot to shot. This structure will ultimately limit the accuracy of our spectroscopic measurements. Published by AIP Publishing.